MED64-Plex 4/8

Microelectrode Array Recording System (AlphaMED)
Description

 

Medium throughput system for recording from 64 electrodes on 4 or 8 samples

The MED64-Plex 4/8 System is a medium throughput MEA system allowing recording at all 64 electrodes from 4 or 8 samples. The MED64’s high fidelity will provide more accurate and reliable data allowing advanced screening.

Features

  • Acquire extracellular signals from 4 or 8 samples (64 electrodes per sample).
  • Stimulation through 1 or to 2 electrodes per sample sequentially.
  • Unrivaled signal-to-noise ratio of 0.8 μV RMS noise level.
  • Large current-driven stimulation via any electrode.
  • Recording inside of a humidified incubator.
  • One system can be used for multiple experiments with its amplifier’s broad acquisition bandwidth and analog lowpass/highpass filter.
  • Analyze data online during your experiment or offline after.
  • Suitable for acute slices, slice cultures, and cell cultures.
  • Low-cost upgrade from MED64-Basic/QuadII/Allegro System.

 

System Configuration

MED Probes     4  (Plex4)      8 (Plex8)

MED Connectors     4  (Plex4)      8 (Plex8)

MED64 Head Amplifier

MED64 Main Amplifier

MED64 Multiplexer I

MED64 Multiplexer II

Mobius Software

Data acquisition PC

 

Specifications

Number of channels / sample
    64

RMS Noise (Typical)
*Short-circuited:
14 nV/√Hz

* MED-515A input:
0.8 μV (f < 3 kHz)
1.2 μV (f < 5 kHz)
1.8 μV (f < 10 kHz)

Number of samples
    4 (or 8)

Gain
    × 200 - 2170

Acquisition bandwidth
    0.1 Hz - 10 kHz

Analog low-cult filter
    0.1, 1, 10, 100 Hz

No. of stimulator channel
    2

Analog high-cut filter
    1.0, 2.0, 2.5, 5.0, 7.5, 10 kHz

Stimulus output format
    Current-driven

Input impedance
    100 MΩ

Max. output current
    ± 200 μA

Output impedance
    NA

Signal output
    Resolution    16 bit
    Output format    USB
    Sampling rate    Up to 20 kHz / channel

 

 

Also See:

PDF file MED64 Product Brochure
( PDF)