MED64-Allegro
The MED64-Allegro System is a most sensitive medium throughput MEA system with multiple wells. Ideal for drug testing using stem cell-derived cardiomyocytes/neurons.
Its high fidelity provides accurate and reliable data by allowing advanced screening as well as analysis during the testing.
Features
- Acquire extracellular signals from 4 or 8 samples (16 electrodes per sample for 4 well / 8 electrodes for 8 well)
- Stimulation through 1 or 2 electrodes per sample sequentially.
- Unrivaled signal-to-noise ratio of 0.8 μV RMS noise level.
- Large current-driven stimulation via any electrode.
- Recording inside of a humidified incubator.
- One system can be used for multiple experiments with its amplifier’s broad acquisition bandwidth and analog lowpass/highpass filter.
- Analyze data online during your experiment or offline after.
- Suitable for cell cultures.
- Low-cost shift to MED64-Quad II/Basic System..
System Configuration
MED Multi-well Probe
MED Multi-well Connector
MED64 Head Amplifier
MED64 Main Amplifier
Mobius Software
Data acquisition PC
Specifications
Number of channels / sample
16 (or 8)
RMS Noise (Typical)
*Short-circuited:
14 nV/√Hz
—
* MED-515A input:
0.8 μV (f < 3 kHz)
1.2 μV (f < 5 kHz)
1.8 μV (f < 10 kHz)
Number of samples
4 (or 8)
Gain
× 200-2170
Acquisition bandwidth
0.1Hz – 10 kHz
Analog low-cult filter
0.1, 1, 10, 100 Hz
No. of stimulator channel
2
Analog high-cut filter
1.0, 2.0, 2.5, 5.0, 7.5, 10 kHz
Stimulus output format
Current-driven
Input impedance
100 MΩ
Max. output current
± 200 μA
Output impedance
NA
Signal output
Resolution 16 bit
Output format USB
Sampling rate Up to 20 kHz / channel
Also See:
MED64 Product Brochure
( PDF)